13. Appendix D Sample Preservation
It is possible to collect samples for later analysis if the situation dictates it. For
this application the YSI 2315 Preservative Tube Kit
YSI 2315 Preservative Tube Kit
YSI 2315 Preservative Tube Kit
YSI 2315 Preservative Tube Kit is available. The kit
consists of re-sealable tubes containing a preservative powder and packets of
YSI 1515 Cell Lysing Agent.
WARNING: You must use the YSI Cell Lysing Agent in your
SPORT when running samples preserved in YSI
preservative tubes.
You cannot use the YSI preservative tubes unless the lysing agent is used in the
instrument when the actual measurement takes place. This is due to the fact that
the preservative tubes contain a detergent. This detergent, when introduced into
a system without the cell lysing agent, can affect the membrane response and
cause incorrect readings. While this detergent is also present in the cell lysing
agent, the response effect is compensated for during normal calibration, and the
sudden introduction of a sample containing the detergent will have no adverse
effect. Refer to Appendix E–Effects Of Selected Substances.
A sample size of 100 microliters minimum is required for each tube.
Samples preserved in YSI preservative tubes can be stored for up to 48 hours @
25°C, and for up to 7 days @ 5°C.
Summary of Contents for 1500 Sport
Page 2: ......
Page 8: ...1 1 1 1 2 2 2 2 ...
Page 46: ...5 5 5 5 1 1 1 12 2 2 2 ...
Page 68: ...6 6 6 6 22 22 22 22 ...
Page 78: ...7 7 7 7 10 10 10 10 ...
Page 86: ......
Page 100: ...9 9 9 9 14 14 14 14 ...
Page 114: ...12 12 12 12 4 4 4 4 ...
Page 116: ...13 13 13 13 2 2 2 2 ...
Page 122: ...14 14 14 14 6 6 6 6 ...
Page 124: ...15 15 15 15 2 2 2 2 ...
Page 126: ...16 16 16 16 2 2 2 2 ...
Page 128: ...17 17 17 17 2 2 2 2 ...
Page 132: ...18 18 18 18 4 4 4 4 ...