JOHNSON CONTROLS
101
FORM 150.62-NM8 (1020)
2
The line under the 0 is the cursor. If the value is wrong,
it may be changed by using the
↑ (UP) and ↓ (DOWN)
arrow keys until correct. Pressing the ENTER/ADV
key will enter the times and then move the cursor to the
minute box. The operation is then repeated if necessary.
This process may be followed until the hour, minutes,
and meridian (AM or PM) of both the START and STOP
points are set. After changing the meridian of the stop
time, pressing the ENTER/ADV key will advance the
schedule to the next day.
Whenever the daily schedule is changed
for Monday, all the other days will
change to the new Monday schedule.
This means if the Monday times are not
applicable for the whole week then the
exceptional days would need to be re
-
programmed to the desired schedule.
To page to a specific day press the SCHEDULE/
ADVANCE DAY key. The start and stop time of each
day may be programmed differently using the ↑ (UP)
and ↓ (DOWN)
arrow, and ENTER/ADV keys.
After SUN (Sunday) schedule appears on the display a
subsequent press of the SCHEDULE/ADVANCE DAY
key will display the holiday schedule. This is a two part
display. The first reads:
H O L S T A R T = 0 0 : 0 0 A M
S T O
P = 0 0 : 0 0 A
M
The times may be set using the same procedure as
described above for the days of the week. After
changing the meridian of the stop time, pressing the
ENTER/ADV key will advance the schedule to the
following display:
S _ M T W T F S
H O L I D A Y N O T E D B Y *
The line below the empty space next to the S is the
cursor and will move to the next empty space when the
ENTER/ADV key is pressed. To set the Holiday, the
cursor is moved to the space following the day of the
week of the holiday and the is pressed. An * will appear
in the space signifying that day as a holiday. The * can
be removed by pressing the DOWN arrow key.
The holiday schedule must be programmed weekly
–once the holiday schedule runs, it will revert to the
normal daily schedule.
TABLE 20 – COOLING SETPOINTS, PROGRAMMABLE LIMITS AND DEFAULTS
SETPOINT KEY
MODE
LOW LIMIT
HIGH LIMIT
DEFAULT
LEAVING CHILLED LIQUID SETPOINT
WATER COOLING
40.0°F
4.4°C
**70.0°F
21.1°C
44.0°F
6.7°C
GLYCOL COOLING
*10.0°F
-12.2°C
**70.0°F
21.1°C
44.0°F
6.7°C
LEAVING CHILLED LIQUID CONTROL RANGE
—
1.5°F
0.8°C
2.5°F
1.4°C
2.0°F
1.1°C
RETURNED CHILLED LIQUID SETPOINT
WATER COOLING
40.0°F
4.4°C
70.0°F
21.1°C
44.0°F
6.7°C
GLYCOL COOLING
10.0°F
-12.2°C
70.0°F
21.1°C
44.0°F
6.7°C
RETURN CHILLED LIQUID CONTROL RANGE
—
4.0°F
2.2°C
20.0°F
11.1°C
10.0°F
5.6°C
MAX EMS-PWM REMOTE TEMPERATURE RESET
—
2°F
1.1°C
40°F
22.2°C
20°F
11.0°C
*
Refer to Engineering Guide for operation below 30°F (-1.1°C)
. Alternate thermal expansion valves must be used below 30°F (-1.1°C).
* When using glycol, Leaving Chilled Liquid Setpoint should not be set below 20°F (-6.7°C).
** Do not exceed 55°F (12.8°C) setpoint before contacting the nearest Johnson Controls
Office for application guidelines.
Summary of Contents for YCAL0014E Series
Page 4: ...JOHNSON CONTROLS 4 FORM 150 62 NM8 1020 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 10: ...JOHNSON CONTROLS 10 FORM 150 62 NM8 1020 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 18: ...JOHNSON CONTROLS 18 FORM 150 62 NM8 1020 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 31: ...JOHNSON CONTROLS 31 FORM 150 62 NM8 1020 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 126: ...JOHNSON CONTROLS 126 FORM 150 62 NM8 1020 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 140: ...JOHNSON CONTROLS 140 FORM 150 62 NM8 1020 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 155: ...JOHNSON CONTROLS 155 FORM 150 62 NM8 1020 5 6 ...
Page 157: ...JOHNSON CONTROLS 157 FORM 150 62 NM8 1020 5 6 LD08855 ...
Page 161: ...JOHNSON CONTROLS 161 FORM 150 62 NM8 1020 161 5 6 ...
Page 163: ...JOHNSON CONTROLS 163 FORM 150 62 NM8 1020 5 6 ...
Page 167: ...JOHNSON CONTROLS 167 FORM 150 62 NM8 1020 5 6 ...
Page 169: ...JOHNSON CONTROLS 169 FORM 150 62 NM8 1020 5 6 ...
Page 171: ...JOHNSON CONTROLS 171 FORM 150 62 NM8 1020 5 6 ...
Page 173: ...JOHNSON CONTROLS 173 FORM 150 62 NM8 1020 5 6 ...
Page 175: ...JOHNSON CONTROLS 175 FORM 150 62 NM8 1020 5 6 ...
Page 177: ...JOHNSON CONTROLS 177 FORM 150 62 NM8 1020 5 6 ...
Page 181: ...JOHNSON CONTROLS 181 FORM 150 62 NM8 1020 5 6 ...
Page 183: ...JOHNSON CONTROLS 183 FORM 150 62 NM8 1020 5 6 LD08833 ...
Page 185: ...JOHNSON CONTROLS 185 FORM 150 62 NM8 1020 5 6 ...
Page 187: ...JOHNSON CONTROLS 187 FORM 150 62 NM8 1020 5 6 ...
Page 189: ...JOHNSON CONTROLS 189 FORM 150 62 NM8 1020 5 6 ...
Page 191: ...JOHNSON CONTROLS 191 FORM 150 62 NM8 1020 5 6 ...
Page 195: ...JOHNSON CONTROLS 195 FORM 150 62 NM8 1020 5 6 ...
Page 197: ...JOHNSON CONTROLS 197 FORM 150 62 NM8 1020 5 6 ...
Page 199: ...JOHNSON CONTROLS 199 FORM 150 62 NM8 1020 5 6 ...
Page 200: ...JOHNSON CONTROLS 200 FORM 150 62 NM8 1020 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 211: ...JOHNSON CONTROLS 211 FORM 150 62 NM8 1020 THIS PAGE INTENTIONALLY LEFT BLANK ...