<6. Components and Their Functions>
6-2
IM 11M13A01-02E
6.1.2
High Temperature Detector (ZR22S-
-015)
Sample gas outlet
When a sample gas pressure is
negative, connect the auxiliary
ejector assembly.
When the sample gas is high
temperature and high pressure,
and does not fall below 700°C,
connect a pressure control valve
(e.g. a needle valve).
Separate type
High Temperature Detector (ZR22S-
-015)
When the temperature of the sample gas is
between 700°C and 1400°C, mount this
detector with a ZO21P probe adapter.
F06-2E.ai
Flange
Selectable from JIS,
ANSI or DIN standards
High Temperature Probe Adapter
(ZO21P)
The probe is made of either SUS310S
or silicon carbide (SiC).
Its length is 0.5, 0.6, 0.7, 0.8, 0.9, 1.0
or 1.5 m.
When using an SiC probe, mount it
vertically downward.
Figure 6.2 High Temperature Detector