<7. Startup>
7-3
IM 11M13A01-02E
7.5
Confirmation of Converter Type Setting
This converter can be used for both the Oxygen Analyzer and the Humidity Analyzer.
Before setting the operating data, be sure to check that the desired converter model has been
set.
Note that if the converter type setting is changed, the operating data that have been set are then
initialized and the default settings remain. To set the desired operating data, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display
(Figure 7.4).
(2) Use the [▼] key to select “Commissioning” and press the [Enter] key.
(3) In the Passwords display, enter the [Enter] key. If the password is to be set again, enter the
new password (for details, see Subsection 8.6.5, “Setting Passwords” later in this manual).
(4) The Commissioning display shown in Figure 7.5 appears. Select “Basic setup” and press
the [Enter] key.
(5) The Basic setup display shown in Figure 7.6 then appears. Confirm the currently set
converter type. If the Humidity Analyzer option /HS was selected at the time of purchase, the
converter was set for high temperature humidity use before shipment.
(6) If the converter type is to be changed, press the [Enter] key. The display shown in Figure 7.7
then appears.
(7) Use the [▼] key to select the type of equipment. Then press the [Enter] key to complete the
converter selection.
(8) If the type of converter is changed after setting the operating data, those data are then
initialized and the default settings remain. Reset the operating data to meet the new type of
equipment.
Enter
Execution/Setup
Blow back
Reset
Setup
Maintenance
Commissioning
Execution
Calibration
F07-4E.ai
Enter
Commissioning
mA-output setup
Alarms setup
Contact setup
Others
Basic setup
F07-5E.ai
Figure 7.4 “Execution/Setup” Display
Figure 7.5 “Commissioning” Display
Enter
Basic setup
This instrument is for
Oxygen
Please note:changing this
Basic setting will return ALL
other setting to Default!!
Detector is ZR22
Select measure gas: Wet
F07-6E.ai
This instrument is for
Please note:changing this
Basic setting will return ALL
other setting to Default!!
Detector is ZR22
Select measure gas: Wet
Enter
Basic setup
Oxygen
Humidity
F07-7E.ai
Figure 7.6 “Basic Setup” Display
Figure 7.7 Equipment Setup
Summary of Contents for ZR22S
Page 11: ...Blank Page...
Page 19: ...Blank Page...
Page 23: ...Blank Page...
Page 52: ...2 Specifications 2 29 IM 11M13A01 02E...
Page 70: ...Blank Page...
Page 134: ...Blank Page...
Page 164: ...Blank Page...
Page 190: ......
Page 192: ......
Page 194: ......
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ...Blank Page...
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......