<10. Other Functions>
10-19
Enter
Open zero gas valve.
Set flow zero gas to
600ml/min.
Valve opened
Cancel calibration
Manual calibration
F10-19E.ai
Figure 10.19 Manual Calibration Display
(2) Adjust the flow rate and select “Valve opened” in the Manual calibration display.
Check the Trend graph display to see that the measured value is stabilized. Then press the
[Enter] key. The Manual calibration display shown in Figure 10.20 appears.
Enter
Zero calibration
Close the zero gas valve.
Span calibration
End
Manual calibration
F10-20E.ai
Figure 10.20 Zero-point Calibration Complete (in Manual Calibration)
(3) Close the zero gas flow setting valve to stop the zero gas flow. If the valve has a lock nut,
be sure to tighten the lock nut to prevent the any leakage of the zero gas into the sensor
because the valve may become loose during measurement.
10.8.4 Operation After Calibration
No special operation of the instrument is needed after calibration. However, it is recommended
that the pressure reducing valve for the zero gas cylinders be closed because calibration is not
required so often.
IM 11M13A01-02E
Summary of Contents for ZR22S
Page 11: ...Blank Page...
Page 19: ...Blank Page...
Page 23: ...Blank Page...
Page 52: ...2 Specifications 2 29 IM 11M13A01 02E...
Page 70: ...Blank Page...
Page 134: ...Blank Page...
Page 164: ...Blank Page...
Page 190: ......
Page 192: ......
Page 194: ......
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ...Blank Page...
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......