![YOKOGAWA ZR22S User Manual Download Page 183](http://html1.mh-extra.com/html/yokogawa/zr22s/zr22s_user-manual_3505092183.webp)
<12. Troubleshooting>
12-7
(4) Check whether the deterioration of or damage to the sensor assembly that caused the
alarm has occurred suddenly during the current calibration in the following procedure:
a. Call up the Detailed-data Display.
b. Use the [Page Scroll] key to check Calibration data (Figure 12.6). The span and zero
correction ratios of the last ten calibration can be checked here. By checking these data,
whether the sensor deterioration has occurred suddenly or gradually can be determined.
Cell voltage:
0 . 6 mV
Thermo voltage:
4 2 . 1 mV
C.J.resistance: 1 1 8 1 . 4 Ω
Cell resistance:
4 4 Ω
Soft.rev.:
0 . 2 4
Tag:
F12-5E.ai
Hold
Calibration data:
1: 2 0 0 0 / 07 / 07 1 4 : 57
Span gas ratio: - 2 . 9 %
Zero gas ratio: 7 0 . 3 %
2: 2 0 0 0 / 07 / 07 1 4 : 5 4
Span gas ratio: - 2 . 9 %
Zero gas ratio: 8 9 . 7 %
Tag:
Hold
F12-6E.ai
Figure 12.5 Detailed-data Display
Figure 12.6 Calibration History
(5) If the sensor assembly has deteriorated suddenly, the check valve that prevents moisture in
the furnace from entering into the calibration pipes may have malfunctioned. If the furnace
gas flows into calibration lines, the gas is cooled and thus condensation develops and
accumulates in the pipe. During calibration the condensate is carried with the calibration gas
and blow onto the sensor assembly, whereby the cell is cooled quickly. This results in the
failed sensor assembly.
(6) If the sensor assembly has deteriorated gradually, check the condition of the sensor
assembly following the procedure below.
a. Use the [Page Scroll] key to check Cell Resistance. It should be 200 Ω or less if the cell
(sensor) is new. On the other hand, if the cell (sensor) is approaching the end of its service
life, it will be 3 to 10 kΩ.
b. Use the [Page Scroll] key to check Cell robustness. It should say “life > 1 year” if the cell
(sensor) is in good condition.
Span gas ratio: - 2 . 9 %
Zero gas ratio: 7 0 . 3 %
Response time: 0 s
Cell robustness: life< 1 month
Cell temperature: 1 0 6 7 °C
C.J.temperature: 4 7 °C
Tag:
Hold
F12-7E.ai
Figure 12.7
IM 11M13A01-02E
Summary of Contents for ZR22S
Page 11: ...Blank Page...
Page 19: ...Blank Page...
Page 23: ...Blank Page...
Page 52: ...2 Specifications 2 29 IM 11M13A01 02E...
Page 70: ...Blank Page...
Page 134: ...Blank Page...
Page 164: ...Blank Page...
Page 190: ......
Page 192: ......
Page 194: ......
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ...Blank Page...
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......