<8. Detailed Data Setting>
8-15
Set period over which
average is calculated:
1 h
Set period over which
maximum and minimum is
stored: 2 4 h
Enter
Averaging
F08-14E.ai
Figure 8.14 Setting Average Value Calculation Periods and
Maximum and Minimum Value Monitoring Periods
Default Values
“Set period over which average is calculated” is set to 1hr, and “Set period over which maximum
and minimum is stored” is set to 24 hrs prior factory shipment or after data initialization.. .
8.6.3
Setting Fuels
Input Parameters
The analyzer calculates the moisture content contained in exhaust gases. The following sets
forth the fuel parameters necessary for calculation and their entries. The moisture quantity may
be mathematically expressed by:
(water vapor caused by combustion and water vapor contained in the exhaust gas)
Moisture quantity =
= (Gw + Gw1)/G x 100
=
where,
Ao : Theoretical amount of air per unit quantity of fuel, m
3
/kg (or m
3
/m
3
) ... 2 in Table 8.9
G: Actual amount of exhaust gas (including water vapor) per unit quantity of fuel, m
3
/kg (or m
3
/m
3
)
Gw : Water vapor contained in exhaust gas per unit quantity of fuel (by hydrogen
and moisture content in fuel), m
3
/kg (or m
3
/m
3
)
............ 1 in Table 8.9
Gw1: Water vapor contained in exhaust gas per unit quantity of fuel (moisture content
in air), m
3
/kg (or m
3
/m
3
)
Go: Theoretical amount of dry exhaust gas per unit quantity of fuel, m
3
/kg (or m
3
/m
3
)
m: Air ratio
X : Fuel coefficient determined depending on low calorific power of fuel, m
3
/kg (or m
3
/m
3
) ....... 3 in Table 8.9
Z : Absolute humidity of the atmosphere, kg/kg ....... Figure 8.16
X + Ao x m
Gw + (1.61 x Z x m x Ao )
Siki08-2E.ai
actual exhaust gas (including water vapor) per fuel
+ (water vapor contained in air for combustion)
x 100
Gw + (1.61 x Z x m x Ao)
Go + Gw + [(m - 1) Ao + (1.61 x Z x m x Ao)]
x 100
x 100
............ Equation 1
............ Equation 2
Fill in the boxes with fuel parameters in Equation 2 above to calculate the moisture content.
Use Ao, Gw and X shown in Table 8.9. If there are no appropriate fuel data in Table 8.9, use the
following equations for calculation. Find the value of “Z” in Equations 1 and 2 using Japanese
Industrial Standards JIS B 8222. If a precise measurement is not required, obtain the value of “Z”
using a graph for the absolute humidity indicated by a dry and wet bulb hygrometer.
IM 11M13A01-02E
Summary of Contents for ZR22S
Page 11: ...Blank Page...
Page 19: ...Blank Page...
Page 23: ...Blank Page...
Page 52: ...2 Specifications 2 29 IM 11M13A01 02E...
Page 70: ...Blank Page...
Page 134: ...Blank Page...
Page 164: ...Blank Page...
Page 190: ......
Page 192: ......
Page 194: ......
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ...Blank Page...
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......