Model YPP6060 Dual-Input Process Meter
Instruction Manual
80
Tare (
tare A
,
tare B
)
The tare function zero’s out the display. In the case of scale weight, tare
is used to eliminate container weight and provide net weight readings.
There are two tare functions; Capture Tare for channel A and B, and
Reset Tare. Display channel indicator letter flashes when a tare is used.
It will flash until the tare is reset.
0.0
lbs
0.2
lbs
0.2
lbs
Before tare
After capture tare
After reset tare
Gross (without tare) and net (with tare) values can be viewed
simultaneously. See page 42.
Internal Calibration (
ICAL
)
The meter is factory calibrated prior to shipment
to read in milliamps and volts depending on the
input selection. The calibration equipment is
certified to NIST standards.
The use of calibrated signal sources is necessary to perform the internal
calibration of the meter.
Check calibration of the meter at least every 12 months. Each input
must be recalibrated separately.
Notes:
1. If meter is in operation and it is intended to
accept only one input type (e.g. 4-20 mA),
recalibration of other input is not necessary.
2. Allow the meter to warm up for at least
15 minutes before performing the internal
calibration procedure.
The
Internal
calibration
menu is part of the
Advanced
Features
menu.
1. Press and hold the Menu button for three seconds to access the
advanced features of the meter.
2. Press the Up
arrow button to scroll to the
Internal calibration
menu
(
ICAL
) and press Enter.
Summary of Contents for YPP6060
Page 73: ...Model YPP6060 Dual Input Process Meter Instruction Manual 73 ...
Page 100: ...Model YPP6060 Dual Input Process Meter Instruction Manual 100 NOTES ...
Page 102: ...Model YPP6060 Dual Input Process Meter Instruction Manual 102 ...
Page 103: ...Model YPP6060 Dual Input Process Meter Instruction Manual 103 NOTES ...