12-26
IM WT3001E-51EN
• Aborting the Initialization
6.
Press the
Reset
soft key. An Alert dialog box appears.
7.
Press the
cursor keys
to select OK.
8.
Press
SET
to abort the initialization. The flicker measurement status indicates
Reset. To continue the initialization, select Cancel in step 7 and press
SET
.
• Starting the Measurement of dmax Caused by Manual Switching
9.
Press the
Start
soft key when the flicker measurement status is Ready.
The measurement of dmax caused by manual switching starts. The flicker
measurement status indicates Start.
10.
Turn the EUT ON to achieve normal operation.
Operate the EUT in the normal condition as long as possible in the measurement
period of one observation period (1 minute).
11.
Turn the EUT OFF before the measurement of one observation period (1 minute)
completes. When the measurement of one observation period is complete, the
flicker measurement status changes to Ready.
12.
Repeat steps 9 to 11 to measure dmax 24 times.
Note
Repeat the measurement after the device that reduces dmax cools to the ambient temperature.
• Holding the Measurement of dmax Caused by Manual Switching
13.
Press
HOLD
. The HOLD key illuminates, and the display is held. The
measurement of dmax caused by manual switching continues.
12.6 Executing the Measurement of dmax Caused by Manual Switching