![YOKOGAWA WT3002E User Manual Download Page 199](http://html1.mh-extra.com/html/yokogawa/wt3002e/wt3002e_user-manual_4132509199.webp)
11-18
IM WT3001E-51EN
• Turning ON/OFF Cursor Measurement
3.
Press the
FFT Cursor
soft key to select ON or OFF.
If ON is selected, the result of the cursor measurement is displayed.
• Selecting the Source FFT for Cursor Measurement
• Selecting the Source FFT for
4.
Press the
FFT C1 Trace
soft key to display the source FFT selection box.
5.
Press the
cursor keys
to select FFT1 or FFT2.
6.
Press the
SET
key to confirm the source FFT.
• Selecting the Source FFT for Cursor x
4.
Press the
FFT C2 Trace
soft key to display the source FFT selection box.
5.
Press the
cursor keys
to select FFT1 or FFT2.
6.
Press the
SET
key to confirm the source FFT.
• Moving the Cursors
7.
Press the
FFT C1
+/
FFT C2
x soft key to set the jog shuttle target to FFT C1 +,
FFT C2 x, or both FFT C1 + and FFT C2 x.
• If you select FFT C1 +, the position of can be moved.
• If you select FFT C2 x, the position of cursor x can be moved.
• If you select both FFT C1 + and FFT C2 x, the two cursors can be moved without
changing the spacing between the two. The value of the digit being specified by FFT
C1 + changes.
8.
Press the
cursor keys
to set the move the cursors.
11.9 Cursor Measurement