![YOKOGAWA WT3002E User Manual Download Page 215](http://html1.mh-extra.com/html/yokogawa/wt3002e/wt3002e_user-manual_4132509215.webp)
12-14
IM WT3001E-51EN
If you set the measurement method to Flicker (normal flicker measurement) in step 2 of
section 12.3, proceed to step 3. If you set the measurement method to dmax
(measurement of dmax caused by manual switching), proceed to step 9, and only set the
judgement condition on maximum relative voltage change dmax (steps 9 to 14).
• Setting the Judgement Conditions for the Relative Steady-State Voltage
Change dc
• Turning ON/OFF the Judgement of Relative Steady-State Voltage Change dc
3.
Press the
cursor keys
to select Judgement by dc.
4.
Press
SET
to select ON (judge) or OFF (not judge).
• Setting the Limit on Relative Steady-State Voltage Change dc
(Valid when ON (judge) is selected in step 4)
5.
Press the
cursor keys
to select Limit by dc.
6.
Press
SET
to display the entry box.
7.
Press the
cursor keys
to set the limit.
8.
Press
SET
or
ESC
to close the box.
• Setting the Judgement Conditions of Maximum Relative Voltage Change
dmax
• Turning ON/OFF the Judgement of Maximum Relative Voltage Change dmax
9.
Press the
cursor keys
to select Judgement by dmax.
10.
Press
SET
to select ON (judge) or OFF (not judge).
• Setting the Limit on Maximum Relative Voltage Change dmax
(Valid when ON (judge) is selected in step 10)
11.
Press the
cursor keys
to select Limit by dmax.
12.
Press
SET
to display the entry box.
13.
Press the
cursor keys
to set the limit.
14.
Press
SET
or
ESC
to close the box.
• Setting the Judgement Conditions for Period during Which Relative
Voltage Change Exceeds the Threshold Level Tmax in a Single Voltage
Fluctuation Period
• Turning ON/OFF the Judgement of Period during Which Relative Voltage
Change Exceeds the Threshold Level Tmax
15.
Press the
cursor keys
to select Judgement by Tmax.
16.
Press
SET
to select ON (judge) or OFF (not judge).
12.4 Setting the Judgement Conditions