12-19
IM WT3001E-51EN
Voltage Fluctuation and Flicker Measurement (Option)
1
2
3
4
5
6
7
8
9
10
11
12
13
Index
• Aborting the Initialization
6.
Press the
Reset
soft key. An Alert dialog box appears.
7.
Press the
cursor keys
to select OK.
8.
Press
SET
to abort the initialization. The flicker measurement status indicates
Reset. To continue the initialization, select Cancel in step 7 and press
SET
.
• Starting the Voltage Fluctuation and Flicker Measurement
9.
Press the
Start
soft key when the flicker measurement status is Ready. The
voltage fluctuation and flicker measurement starts. The flicker measurement
status indicates Start.
• Holding the Voltage Fluctuation and Flicker Measurement
10.
Press
HOLD
. The HOLD key illuminates, and the display is held. The voltage
fluctuation and flicker measurement continues.
• Releasing the Hold Operation
11.
Press
HOLD
while the display is held. The HOLD key turns OFF, and the display
is updated. In the hold condition, the display can be updated to the most recent
value by carrying out single measurements (by pressing SINGLE).
• Aborting the Voltage Fluctuation and Flicker Measurement
12.
Press the
Reset
soft key. An Alert dialog box appears.
13.
Press the
cursor keys
to select OK.
14.
Press
SET
. The voltage fluctuation and flicker measurement is aborted. The
measured data and judgement up to that point are discarded. The flicker
measurement status indicates Reset. To continue the voltage fluctuation and
flicker measurement, select Cancel in step 13 and press
SET
.
12.5 Executing the Normal Voltage Fluctuation and Flicker Measurement