10-7
IM WT3001E-51EN
W
aveform Computation (Option)
1
2
3
4
5
6
7
8
9
10
11
12
13
Index
Explanation
You can create two equations (Math1 and Math2).
• Setting the Equation
• Operand
The following operands can be combined to create an equation. There can be up to
16 operands in an equation.
Operand
Description
U1 to U4
Voltage waveform of each input element
I1 to I4
Current waveform of each input element
P1 to P4
Voltage waveform × current waveform of each input element (instantaneous
power waveform)
TORQUE
Torque signal waveform of the motor input
SPEED
Speed signal waveform of the motor input
Pmi
TORQUE × SPEED (instantaneous waveform of motor output)
K1 to K8
Constant
• Operator
The following operators can be combined to create an equation.
Operator
Setting Example
Description
+, –,
*
, /
U1+I1
Four arithmetic operation of the specified waveform.
ABS
ABS(U1)
Absolute value of the specified waveform
SQR
SQR(U1)
Square of the specified waveform
SQRT
SQRT(U1)
Square root of the specified waveform
LOG
LOG(U1)
Natural logarithm of the specified waveform
LOG10
LOG10(U1)
Common logarithm of the specified waveform
EXP
EXP(U1)
Exponent of the specified waveform
NEG
NEG(U1)
Negation of the specified waveform
AVG2
AVG2(U1
*
I1)
Average of the specified waveform with an average
constant of 2
AVG4
AVG4(U1
*
I1)
Average of the specified waveform with an average
constant of 4
AVG8
AVG8(U1
*
I1)
Average of the specified waveform with an average
constant of 8
AVG16
AVG16(U1
*
I1)
Average of the specified waveform with an average
constant of 16
AVG32
AVG32(U1
*
I1)
Average of the specified waveform with an average
constant of 32
AVG64
AVG64(U1
*
I1)
Average of the specified waveform with an average
constant of 64
• Number and Type of Characters That Can Be Used in the Equation
• Number of Characters
50 characters or less
• Types of Characters
Characters that are displayed on the keyboard and spaces
• Setting the Constants
The values can be set in the range of –9.9999E+30 to 9.9999E+30.
10.4 Setting the Waveform Equation