<8. Maintenance Schedule>
8-6
IM 12Y18A04-01EN
1st Edition : Oct. 04, 2019-00
8.5
Installing Tubing Kit
The tubing in the WE440 Phosphate Analyzer consists of two parts, the Main Tubing Assembly
(K9705CK) and the Pump Head Tubing Assembly (K9705CL). The two sections of tubing are
connected via luer fittings. The complete tubing layout is shown in below.
STEP 1:
lnstall the tubing channel using the pictures below. Note the the white labels will need to
be on the left side of the analyzer. The tubing channel will connect into the upper receiving
channel.
STEP 2:
Insert vent tube 1 and 2 through vent holes in the bottom left of the enclosure as shown in
the picture.
Summary of Contents for WE440
Page 1: ...User s Manual Model WE440 Phosphate Analyzer IM 12Y18A04 01EN IM 12Y18A04 01EN 1st Edition ...
Page 9: ...Blank Page ...
Page 25: ...Blank Page ...
Page 33: ...Blank Page ...
Page 41: ...Blank Page ...
Page 63: ...Blank Page ...
Page 67: ...Blank Page ...