<1. General Information>
1-3
IM 12Y18A04-01EN
1st Edition : Oct. 04, 2019-00
Important standard analyzer capabilities include:
• Sampling the process stream automatically.
• Calculating concentration of your sample and reporting it in units of concentration.
• Recalibrating automatically at preprogrammed intervals to compensate for any changes in
captured sample volume, pump flow rates and reagent composition.
• Cleaning automatically (when programmed) at user-specified intervals to prevent
precipitation build-up, coating of the optics or bacterial/fungal growth.
The analyzer brings colorimetric laboratory precision to the real world of unattended process
analyzing. It is designed for long life under continuous, rigorous use.
The WE400 series analyzers are designed for accommodating several applications. Each
application requires a specific configuration in hardware and software that is set at the factory.
This analyzer model is intended to use for clean water applications such as power plant water
analysis only, suspended solids must be less than 60 microns.
1.2.1
User Interface
n
General Description
The user-interface consists of a keypad and a graphical LCD display. The keypad is located
to the right of the display. The “F1-F4” function keys (soft-keys) take on different meanings
depending on the menu selection. The column on the left of the soft-keys labels the keys with
their appropriate functions.
Keys
RUN:
starts the analysis
STOP:
stops the current functioning of the instrument.
+ , -:
allows for navigation between pages
ENTER: confirms selections made in the menus
MENU: accesses the menus, Pressing the MENU key displays the instrument’s Main
Menu. System parameters can be configured from the Main Menu.
soft-keys (in appropriate menus):
additional navigations for left / right or up / down cursor movement.
LCD display
the date, analysis label, time, navigational menus, current user, instrument-working mode,
current status of the instrument and the soft-key labels.
PHOSPHATE
+SETUP
+MEASUREMENT
+VIEW DATA AND ALARMS
+SERVICES
06/06/19
MAIN MENU
02:20 PM
Data field Cursor
Application name field
Display field Time field
Function keys assignment
Function keys
Operational keys
Figure 1.2