<7. Terms Definition>
7-2
IM 12Y18A04-01EN
1st Edition : Oct. 04, 2019-00
Terms
Description
Possible Values, Options
ANALYSIS PARAMETERS
Mode
Instrument RESTART mode.
Auto Restart, Manual Restart
Measuring Frequency
In the RUN mode, the frequency is the
time duration between two analysis
cycles.
15, 30, 45 minutes, 1 hr, 0.20 hr are
selectable and editable minutes range
00.00 to 99.99Hr
Measuring Units
Units for the analyzed sample.
Six characters long. Alphabets, numerals
and special symbols are allowed.
Analysis Label
Six characters long. Alphabets, Numerals
and +, - are allowed.
Cell Wash Flow Rate
The rate at which the sample is pumped
to wash the cell.
Depends on the inner diameter of the
pump tubing. It is derived from the
calibration of pumps.
Reagent 1–3 Flow Rate The rate at which the reagent is pumped
to wash the cell.
Depends on the inner diameter of the
pump tubing. It is derived from the
calibration of pumps.
Stability Criteria
Defines the stability level of the data
before accepting it. The stability criteria is
a measure of stability of the absorbance
level rate of 0.001 AU per sec as
specified by the number 1 thrugh 10.
1 thrugh 10.
Post Wash Event
Washing the cell after completion of the
event selected with the following source,
volume and stir time.
Analysis, Calibration, Validation, Not
Assigned.
Post Wash Cycle Source Fluid for post washing the reaction cell.
Sample, Reagent, Cleaner, Standard
Post Wash Cycle
Volume
The amount of fluid pumped for post
washing.
-99.9 to +99.9 mL
Post Wash Cycle Stir
Time
The amount of time the fluid is stirred for
post washing the cell.
0 to 999 seconds
CLEANER PARAMETERS
Cleaner Source
Fluid for cleaning the cell.
Sample, Reagent, Cleaner, Standard,
Not Assigned
Volume to add
The amount pumped for cleaning.
0 to 255 mL
Clean Flow Rate
The rate at which the cleaner source is
pumped to clean the cell.
Depends on the inner diameter of the
pump tubing. It is derived from the
calibration of pumps.
Stir Time
The amount of time the fluid is stirred for
cleaning the cell.
0 to 999 seconds
Frequency
The time duration between two cleaning
cycles.
Manual, Daily, Weekly, Monthly
Start Day/Date
For weekly cleaning the day of cleaning.
For monthly cleaning the date of
cleaning.
Day - Week days
Date - Current month dates (1 to 31)
Start Time
The time to start the cleaning on the day
or date selected.
00:00 hr to 23:59 hr
TEMPERATURE UNITS
Units
Units of temperature.
Celsius, Fahrenheit
METHOD PARAMETERS
Cell Wash Volume
The amount of sample pumped to wash
the cell.
1 to 100 mL
Reaction Temperature
Control
PID temperature control.
Enable, Disable
Reaction Temperature
The temperature set point.
0 to 60
Wavelength
The LED wavelength.
Blank Measurement
Allows the blank measurement in the
analysis cycle.
Disable, Before Reagent
Reagent Volume
The amount of reagent pumped in to the
cell.
0 to 6.00 mL
Stir Time
The amount of time the fluid is stirred
after pumping the reagent.
0 to 999 seconds