6. Menu Overview
Explanations of the four major menu sections are described as follows:
Setup - For configuring the instrument’s functionality and operational parameters, such as
measurement range, measurement frequency, display contrast, date, time, and more.
Measurement - For performing the manual analysis, calibration, validation and cleaning cycles.
View Data and Alarms - For reviewing the collected data of up to the last 31 days in text and
graphics mode, and for reviewing and resetting the various Alarms and Faults triggered
during the latest cycle performed.
Services - For system administration, hardware setup, prime fluidics and diagnostics; and
control of the various devices including LED, pumps, input level sensors, relays and analog
outputs.
NOTE
Press the STOP key to exit Auto RUN mode to access the Service Menu.
6.1
Setup Menu
Alarms Menu
Allows configuration of alarm parameters (type, value and hysteresis). Also allows
assignment of individual alarms to faults.
Relays Mapping Menu
Allows mapping of multiple alarms to individual relays.
Analog Outputs Menu
Allows configuration of individual analog outputs.
Inputs Menu
Allows configuration of five internal inputs (for fluidics sensors) and two external inputs (for
instrument control).
Display Parameters Menu
Allows configuration of display-related parameters.
Calibration Parameters Menu
Allows configuration of parameters for calibration. The menu shows two columns tagged as
REFERENCE and CURRENT. The parameters listed under the REFERENCE column are
set for validating auto calibrations repeats. The new calibration results are tested against the
reference calibration parameters listed under the REFERENCE column. The CURRENT
column shows the current operating parameters that are being used by the instrument when
calculating the analysis measurement result. Calibrations are similar to analysis cycles and
are on the method set in METHOD PARAMETERS menu.
Validation Parameters Menu
Allows configuration of parameters for validation, choice of validation source, and repetitions
required.
Analysis Parameters Menu
Allows configuration of parameters for instrument mode, measurement frequency,
measurement units, analysis label, cell wash and reagent flow rates, stability criteria and
post wash events. Separate post wash events can be programmed for analysis, calibration,
and validation cycles. All post wash cycles can be turned OFF when set to the “NOT
ASSIGNED”.
Cleaner Parameters Menu
Allows configuration of the parameters for manual or periodic cleaning of the reaction cell.