<2. SPECIFICATIONS>
2-4
IM 12E04A02-02E
Error Detection Function: Turbidity overrange, lamp disconnection, lamp voltage failure, AD
circuit failure, memory failure and CPU failure.
Check Function:
Converter operation check
Manual Calibration:
Zero calibration:
Zero turbidity water or light source off zero, selectable
Span calibration:
Calibration plate
Other Functions
Bi-linear output, high/low alarm and average coef
fi
cient setting
Material:
Detector: Modi
fi
ed PPE, black (wetted part)
Piping:
Rigid PVC, polyethylene resin, and polypropylene resin (wetted
parts)
Stanchion:
Carbon steel plate or stainless steel (optional)
Converter:
Aluminum alloy casting
Finish and Color:
Converter:
Polyurethane resin, baked
fi
nish, Munsell 0.6GY3.1/2.0 and Munsell
2.5Y8.4/1.2
Stanchion:
Polyurethane resin, baked
fi
nish, Munsell 0.6GY3.1/2.0
Ambient Temperature: -5 to 50 °C (When water sample and tap water freeze in winter, a
protection against freeze is required.)
Ambient Humidity:
5 to 95%RH, non-condensing
Storage Temperature: -30 to 70 °C
Weight:
Detector: Approx. 3 kg
Converter: Approx. 9 kg
With sampling system: Approx. 50 kg (in full speci
fi
cations)
Dimensions:
Detector: 245 (W) x 250 (H) x 200 (D) mm
Converter: 260 (W) x 340 (H) x 150 (D) mm
With stanchion: 530 (W) x 1450 (H) x 550 (D) mm
2.2 Characteristics
Linearity (when using kaolin standard):
Upper range limit of 1000 mg/l (degrees) or less: ±2% of full scale
Upper range limit of 2000 mg/l (degrees) or less: ±5% of full scale
When the lower range value is a value other than 0 mg/l (degree),
the full scale should be interpreted as the upper range limit.
Linearity (when using polystyrene latex standard):
Upper range limit of 100 mg/l or less:
±2% of full scale
Upper range limit of 2000 mg/l or less:
±5% of full scale
Repeatability:
Standard solution:
2% of full scale
Calibration plate:
1% of full scale
Effect of Power Supply Voltage: Within ±1% of full scale/±10% of rated voltage
Effect of Ambient Temperature: ±1% of full scale/10 °C (calibration plate)
Response Time:
Within 2 minutes (90% response, with sampling system, at a
fl
ow rate of 3
l/min)
Warm-up Time:
Approx. 30 minutes
3rd Edition : May. 31, 2010-00