<8. MAINTENANCE>
8-10
IM 12B01B30-02EN
1st Edition : Mar. 08, 2023-00
n
Replacement of Fuses
Be sure to use a A1633EF for the fuse of TB830D.
(1) Loosen the four (4) setscrews to open the detector top cover.
(2) Take out the fuse. (Figure 8.10).
(3) Replace the fuse with a new one.
(4) Put the top cover back in place.
Fuse
Figure 8.10
Fuse replacement
n
Replacing the fuse in the relay box (when -A2, -A3, or -A5 is selected)
When Sampling system: -A2, -A3, or-A5 (with auto cleaning) is specified, a relay box is included.
Be sure to use A1624EF for the fuse of the relay box.
(1) Loosen the four (4) screws and open the relay-box window.
(2) Remove the fuse holders (2 pcs.) (Figure 8.11).
(3) Replace the fuse with a new one and install the fuse holder.
(4) Close window of the relay box.
Fuse
Figure 8.11
Fuse replacement (relay box)