<8. MAINTENANCE>
8-6
IM 12B01B30-02EN
1st Edition : Mar. 08, 2023-00
CAUTION
Do not mix up two different sizes (1 μm and 0.2 μm) of the zero turbidity filter element if you install
two zero turbidity filters.
Air vent
O-ring
Filter element
Remove
Case
Figure 8.5
Zero Turbidity Filter Element Replacement
Table 8.3
Filter element
Part No.
Name
K9008ZD
1 μm filter element
K9726EH
0.2 μm filter element
K9411UD
O-ring
8.8
LED Replacement
CAUTION
• Turn off the power of TB830D before replacing the LED.
• LED may be damaged by static electricity. When replacing the LED be sure to take
measures against static electricity, such as wearing a grounded wristband.
LED light source usually lasts more than 3 years. However, it is recommended to replace it every
3 years for preventive maintenance.
This section explains how to replace the LED.
(1) Stops supplying Power Supply to TB830D.
(2) Remove the top cover of the detection unit by loosening the four (4) setscrews.
(3) Loosen the four (4) setscrews on LED ASSY.
(4) Remove the cable connector from LED ASSY (Figure 8.7).
(5) Replace LED ASSY with a new one.
(6) Attach the cable connector to the new LED ASSY (Figure 8.8).