<5. FLXA402T sensor menu>
5-6
1st Edition : Mar. 08, 2023-00
IM 12E04B40-02EN
5.6
Maintenance
Sensor menu > Maintenance
Figure 5.8
When you have a password, the password verification appears. Upon successful confirmation,
Maintenance screen appears.
For further information on the password setting, see FLXA402T Operation of Converter IM
12A01G01-03EN 5.4 Password.
5.6.1
Maintenance mode
Switches the mode between the measurement and the maintenance. For Maintenance mode,
See “3. PREPARATION for OPERATION”.
5.6.2
Light source On/Off
Switches On/Off Light source. This function is enabled only when Maintenance mode is On.
5.6.3
SV/Wiper operation
SV/Wiper operation
TB830D does not use wipers. The function controls the solenoid valve connected to Relay box.
You can go to the same screen when tapping (MAN SW
of Main menu. This function is
enabled only when Maintenance mode is On.
CAUTION
SV4 (pinch valve) is set to automatically turn off after 10 minutes of operation to prevent
overheating due to energization.