Setup Operations for Convenient Functions (Setting Mode)
3
3.11 Performing Calibration Correction (/CC1 Option)
The scaled values of the measuring input signal are corrected using specified segments,
and the results are used as measured values.
If you change the input range, set the measured and correction values again.
Procedure
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Carry out the procedure shown in the figure below.
Press the
or
key to select the value.
For the procedure on how to enter values or characters, see page 19 in the
Operation Guide
.
To change the polarity, press the
or
key when the leftmost digit of the
value is blinking.
If you press the
key, the operation is cancelled, and the display returns to a
higher level menu.
(CALIB)
(CH1)
(ON)
(STORE)
(ON)
Select ON to set the calibration correction.
Select the channel number.
Enter the correction value of the 1
st
point.
(Displays the following when the
calibration correction is ON)
Enter the measured value of the 1
st
point.
Select “CALIB.”
Enter the measured value of the 2
nd
point.
Enter the correction value of the 2
nd
point.
Channel number
Measured value of the
1
st
point
Measured value of the
2
nd
point
Enter the measured value of the 16
th
point.
Measured value of the
16
th
point
Enter the correction value of the 16
th
point.
Save the settings.
The settings are activated.
Correction value of the
16
th
point
Correction value of the
1
st
point
Correction value of the
2
nd
point
:
:
Setting example when the calibration correction
method is set to revise value and the number of
calibration points is set to 16