Quick Reference Instruction Manual for Spare
Sensor replacement
Installation of the Spare sensor
18 / 76
IM01U10A01-00EN-R, 1
st
edition, 2019-12-09
6.2.3 Sensor installation
DANGER
Risk of injury due to escaping fluids, if pipe connection is faulty
▶
Correct slope and mismatch of pipe connections before inserting the sensor.
Fig. 5:
Avoid: Slope and mismatch
▶
Avoid fixing anything directly to the sensor. Doing so may result in increased devia-
tions.
1
1
2
Fig. 6:
Installation to be avoided: Fixing the sensor
1
1
2
Fig. 7:
Recommended installation: use the piping to support the sensor
1
Pipe
2
Sensor
▶
Secure pipes before installing the flow meter.
▶
Avoid damaging the process connections.
▶
Flush new pipes before installing the flow meter to remove foreign matter, such as
shavings or other residues.