Quick Reference Instruction Manual for Spare
Transmitter replacement
Parameter setting
44 / 76
IM01U10A01-00EN-R, 1
st
edition, 2019-12-09
7.3.3 Set up process variable units
For Spare
transmitter:
Set the Spare transmitter measurement units to the values from the file “Customer_Set-
tings_YourDeviceSerialNumber.csv” found on the DVD of Spare transmitter.
For Spare sensor:
Set the existing transmitter measurement units to the values from the file “Customer_Set-
tings_YourDeviceSerialNumber.csv” found on the DVD of Spare sensor.
EXAMPLE, see
table below
:
MASS_FLOW
MASS flow unit R/W3
5
0:g/s
1:g/min
2:g/h
3:kg/s
4:kg/min
5:kg/h
6:kg/d
7:t/min
8:t/h
9:t/d
10:lb/s
11:lb/min
12:lb/h
13:lb/d
14:User unit
For this device, configured for mass flow measurement, all relevant parameters like
Qnom, Qmax are saved in kg/h in the factory-backup file.
To change the mass flow unit over the display, please execute the following sequence:
1. Set user level “Specialist”
2. Go to menu item “[Easy setup wizard] ► [Std dev var] ► [Mass] ► [Unit]”
3. Choose unit corresponding to the saved value in the Customer_Settings file, for ex-
ample “[kg/h]”
4. Set chosen Unit “[kg/h]”
5. Check process unit of the exchange transmitter
This parameter is also available in additional menus. The values can be set in each of
these menus:
▪ Basic setup
▪ Detailed setup
Execute this procedure for all other relevant process parameter units in your device cor-
responding described in the
IM 01U10S0
␣
-00
␣␣
-R
.
WARNING
If the units for the measurement parameters aren’t adjusted the measure-
ment will be faulty and the transmitter may show configuration error.
Please note your user settings of the units down before you change it. After the pairing
process you should set the units back to your origin settings!