![YOKOGAWA ROTAMASS Total Insight User Manual Download Page 62](http://html.mh-extra.com/html/yokogawa/rotamass-total-insight/rotamass-total-insight_user-manual_911337062.webp)
Quick Reference Instruction Manual for Spare
Transmitter replacement
Wiring
62 / 76
IM01U10A01-00EN-R, 1
st
edition, 2019-12-09
Active pulse
output
P/Sout
with internal
pull-up resistor
Value
Internal power supply
24 V
DC
±20 %
Internal pull-up resistor
2.2 kΩ
Maximum pulse rate
10000 pulses/s
Frequency range
0 – 12.5 kHz
P/Sout+
P/Sout-
24 V
0 V
ROTAMASS
1
Fig. 37:
Active pulse output
P/Sout
with internal pull-up resistor
①
Electronic counter
Passive pulse
output
Value
Maximum load current
≤ 200 mA
Power supply
≤ 30 V
DC
Maximum pulse rate
10000 pulses/s
Frequency range
0 – 12.5 kHz
ROTAMASS
Pulse+
Pulse-
3
2
1
Fig. 38:
Passive pulse output connection with electronic counter
①
Passive pulse
②
Load resistance
③
Electronic counter
ROTAMASS
Pulse+
Pulse-
3
2
1
Fig. 39:
Passive pulse output connection with electromechanical counter
①
Passive pulse
②
Protective diode
③
Electromechanical counter