<8. EXPLOSION-PROTECTED TYPE INSTRUMENTS>
8-8
IM 01R01B30-00E-E 6th edition October 01, 2014 -00
All Rights Reserved. Copyright © 2003, Rota Yokogawa
i
i
i
i
i
-
+
+
-
o
o
o
o
o
-
+
-
+
~
~
Limit switch SC2-NO type 2
I = 25mA
P = 64mW
L = 0.15mH
C = 150nF
IECEx DEK 12.0003X
Ex ia IIC T6...T4 Gb
U = 16V
EN 60947-5-6 (NAMUR)
Limit switch
RAKD - T80
IECEx PTB 11.0031X
L = 97mH
C = 2320nF
I = 19.1mA
P = 51mW
Transmitter relay
KFA6-SR2-Ex2.W
U = 10.6V
[Ex ia] IIC
KFA6-SR2-Ex2.W
Power supply 230V AC
Limit switch MIN
Limit switch MAX
BU
2
BN
BU
BN
1
2
1
10
12
6
4
3
1
11
7
8
9
14
15
MIN
F
MAX
S
T
U
Hazardous Area
Safe Area
F4.EPS
Fig. 8-3
Ex-Version acc. IECEx (Option /ES1) with 2 limit switches and transmitter relay
Limit switch / Puls
Ex ia IIC T6...T4 Gb
U = 16V
P = 64mW
I = 20mA
C = 0nF
L = 0mH
Puls output
Ex ia IIC T6...T4 Gb
U = 10.6V
L = 0.15mH
C = 150nF
U = 16V
I = 25mA
P = 64mW
L = 97mH
C = 2320nF
I = 19.1mA
P = 51mW
Limit switch
SC2-NO type 2
EN 60947-5-6 (NAMUR)
Limit switch or
Pulse output
KFA6-SR2-Ex1.W
KFA6-SR2-Ex1.W
Transmitter relay
[EEx ia] IIC
4-20mA output
I = 100mA
U = 30V
L = 0.73mH
C = 2.4nF
P = 0.75W
Ex ia IIC T6...T4 Gb
RAKD-E80 Option /ES1
Transmitter power supply
certified acc. IECEx
Output
4-20mA
Power supply
230V AC
i
i
i
i
i
o
i
i
i
i
i
o
o
o
o
-
P
+
P
-
-
+
~
~
i
i
i
i
i
+
L-
L+
-
+
-
+
BU
1 K
15 K
2
1
2
3
1
14
7
8
9
15
BN
1
1
2
+/~
1
IN
-/~
5
6
IN
10
OUT
4
OUT
9
S
S
U
F
G
T
T
Hazardous Area
Safe Area
IECEx DEK 12.0003X
IECEx DEK 12.0003X
IECEx PTB 11.0031X
IECEx DEK 12.0003X
Fig. 8-4
Ex-Version acc. IECEx (Option /ES1) with electronic transmitter in combination with
power supply and additional limit switch or pulse output with transmitter relay
8.3.2 Installation