IM 01R01B30-00E-E 6th edition October 01, 2014 -00
<7. TECHNICAL DATA>
7-10
All Rights Reserved. Copyright © 2003, Rota Yokogawa
RAKD with EAC- certification “intrinsic safe”
(Russia, Belorussia, Kazakhstan) (option /GS1)
Certificate :
RU С-DЕ.ГБО8.В.00317
Output signal :
4–20 mA / Pulse output / Limit switches
Explosion proof :
Ex ia IIC T6 ... T4 Gb
Entity parameter :
Table 12
Analog
output
Pulse
output
Limit
switch
type 2
/K1-/K3
Limit
switch
type 3
/K1-/K3
Ui [V]
30
16
16
16
Ii[ mA]
100
20
25
52
Pi [mW]
750
64
64
169
Li [mH]
0.73
0
0.15
0.15
Ci [nF]
2.4
0
150
150
Temperature specification
:
Table 13
Configura-
tion
Max.
ambient
temperature
Max.
process tem-
perature
Temperature
class
Indicator
“E” + limit
switches type
“2”
65°C
65°C
T6
50°C
80°C
45°C
100°C
T5
38°C
135°C
T4
Indicator “T”
with limit
switches type
“2”
65°C
65°C
T6
80°C
80°C
T5
59°C
100°C
100°C
100°C
T4
73°C
135°C
Indicator “T”
with limit
switches type
“3”
24°C
65°C
T6
37°C
80°C
T5
34°C
100°C
57°C
80°C
T4
54°C
100°C
48°C
135°C
RAKD with IECEx- certification “intrinsic safe ic”
(option /ES3)
Output signal :
4–20 mA / Pulse output / Limit switches
Explosion proof :
Ex ic IIC T6; Gc group II ; category 3G
Entity parameter :
see table 8
Temperature specification
:
see table 9
RAKD with NEPSI- certification “intrinsic safe”
(China) (option /NS1) :
Certificate :
GYJ101552
Output signal :
4–20 mA
Explosion proof :
Ex ia IIC T5/T6
Max. Tamb. :
65°C
Limit switches :
option /K1 to /K8, see certificates GYJ11.1505X
Entity parameter :
Table 10
Analog
output
Pulse
output
Limit
switch
type 2
/K1-/K3
Limit
switch
type 3
/K1-/K3
Limit
switch
type 2
/K6-/K8
Limit
switch
type 3
/K6-/K8
Ui [V]
30
16
16
16
16
16
Ii[ mA]
100
20
25
52
25
52
Pi [mW]
750
64
64
169
64
169
Li [mH]
0.73
0
0.15
0.15
0.1
0.1
Ci [nF]
2.4
0
150
150
30
30
Temperature specification
:
Table 11
Max. ambient
temperature
Max. process
temperature
Temperature
class
65°C
65°C
T6
50°C
80°C
T6
45°C
95°C
T5
Description of limit switch type 2 and 3 see ATEX certificates
from Pepperl & Fuchs:
- PTB 99 ATEX 2219X ( SC2-NO) for /K1 to /K3
- PTB 00 ATEX 2049X (SJ2-S.N) for /K6 to /K10
RAKD with PESO- certification (India)
Option /KS1 must be selected. PESO- certificate is available
at your Yokogawa Sales Office.
RAKD with KOSHA- certification (Korea)
Option /ES1 must be selected.
Same data as for IECEx certification.
Summary of Contents for RAKD series
Page 2: ...Blank Page ...