3-18
IM 04L57B01-01EN
For bar graph display examples, see the following manual.
GX/GP User’s Manual
Section 1.2.3, “Setting the Display”
GM User’s Manual
Section 2.3.3, “Setting the Display”
Division
Set the number of divisions to make with the main scale marks.
Partial (GX/GP only)
*
Setup Item
Selectable Range or Options
Default Value
On/Off
Off/On
Off
Position
1 to 99(%)
50%
Boundary
Span lower limit + 1 digit to span upper limit – 1
digit
0.01
* Appears when in the
Display settings
of the setting menu, the trend partial expansion
On/Off
is set to
On
.
On/Off
Set this to
On
to enable partial expanded display of waveforms.
For details on the function, see the following manual.
GX/GP User’s Manual
Section 1.10.4, “Setting Trend Display Conditions”
Position
Set at which position to display the value specified by
Boundary
within the display width.
Specify a percentage.
Boundary
Set the value that is to be the boundary between the reduced section and the expanded
section in the range of “minimum span value + +1 digit to maximum span value – 1 digit.”
For channels that are set to linear scaling, the selectable range is “minimum scale value +
+1 digit to maximum scale value – 1 digit.”
Example: Span: 0 to 100. Expand: 30. Boundary: 50
The 0 to 50 range is displayed in the 0% to 30% range, and the 50 to 100 range is
displayed in the 30% to 100% range.
Color scale band
Setup Item
Selectable Range or Options
Default Value
Band area
Off, In, Out
Off
Color
24 colors (red, green, blue, blue violet, brown,
orange, yellow-green, light blue, violet, gray,
lime, cyan, dark blue, yellow, light gray, purple,
black, pink, light brown, light green, dark gray,
olive, dark cyan, and spring green) and a user-
defined color (1 color)
—
Display position Lower
Span (scale) lower limit to span (scale) upper
limit
0.00
Display position Upper
Span (scale) lower limit to span (scale) upper
limit
1.00
Band area
Displays a specified section of the measurement range using a color band on the scale. This
setting is shared with the bar graph display.
Options
Description
Off
Disables the function.
In
Displays the area inside using the color band.
Out
Displays the area outside using the color band.
Color
Set the display color.
Upper display position limit, Lower display position limit
Set the display position. Set a value within the span range.