Table 9.5.5 Range of Measurable Air Flow Rate at Selected Process Pressures
min.
max.
min.
max.
min.
max.
min.
max.
min.
max.
min.
max.
min.
max.
min.
max.
min.
max.
min.
max.
DY015
DY025
DY040
DY050
DY080
DY100
DY150
DY200
DY250
DY300
DY025
-/R1
DY040
-/R1
DY050
-/R1
DY080
-/R1
DY100
-/R1
DY150
-/R1
DY200
-/R1
—
—
—
DY040
-/R2
DY050
-/R2
DY080
-/R2
DY100
-/R2
DY150
-/R2
DY200
-/R2
—
—
—
—
Model Code
Flow
Rate
Limits
Minimum and Maximum Measurable Flow Rate in Nm
3
/h
0 MPa
4.8(11.1)
48.2
11.0(19.5)
149
21.8(30.0)
356
36.2(38.7)
591
70.1
1140
122
1990
268
4358
575
7792
1037
12049
1485
17256
0.1 MPa
6.7(11.1)
95.8
15.5(19.5)
297
30.8
708
51
1174
98.4
2266
172
3954
377
8659
809
15482
1461
23939
2093
34286
0.2 MPa
8.2(11.1)
143
19.0(19.5)
444
37.8
1060
62.4
1757
120
3391
211
5919
485
12960
990
23172
1788
35833
2561
51317
0.4 MPa
10.5(11.1)
239
24.5
739
48.7
1764
80.5
2922
155
5642
272
9847
808
21559
1445
38549
2306
59611
3303
85370
0.6 MPa
12.5
334
29.0
1034
61.6
2468
102
4088
197
7892
334
13775
1131
30163
2202
53933
3127
83400
4479
119441
0.8 MPa
16.1
429
33.3
1329
79.2
3171
131
5254
254
10143
442
17703
1453
38765
2599
69313
4019
107181
5756
153499
1 MPa
19.7
524
40.6
1624
97
3875
161
6420
310
12394
540
21632
1776
47365
3175
84693
4911
130968
7033
187556
1.5 MPa
28.6
762
59.0
2361
149
5634
233
9335
451
18021
786
31453
2583
68867
4617
123138
7140
190418
10226
272699
2 MPa
37.5
1000
77.5
3098
184
7394
306
12249
591
23648
1031
41274
3389
90373
6059
161591
9370
249881
13419
357856
2.5 MPa
46.4
1238
95.9
3836
229
9153
379
15164
732
29274
1277
51095
4196
111875
7501
200046
11600
309334
16612
443017
(1) At standard conditions STP (0˚C. 1atm).
(2) Pressure listed is at process temperature of 0˚C.
(3) Maximum flow rate is the lower of 80 m/s.
(4) Minimum values are determined from Table 12. The values in parenthesis show the minimum linear flow rates (Re = 20,000 or 40,000)
when they are higher than the minimum measurable flow rate.
DYF Tab-15
Table 9.5.6 Range of Measurable Saturated Steam Flow Rate at Selected Process Pressures
Minimum and Maximum Measurable Flow Rate in kg/h
0.1 MPa
5.8(10.7)
55.8
13.4(18.9)
169.7
26.5(29.2)
405
44.0
671
84.9
1295
148
2261
324
4950
697
8851
1256
13687
1799
19602
0.2 MPa
7.0(11.1)
80
16.2(20.0)
247.7
32
591
53
979
103
1891
179
3300
392
7226
841
12918
1518
19977
2174
28609
0.4 MPa
8.8(11.6)
129
20.5
400
40.6
954
67.3
1580
130
3050
227
5326
498
11661
1068
20850
1929
32243
2762
46175
0.6 MPa
10.4(12.1)
177
24.1
548
47.7
1310
79
2170
152
4188
267
7310
600
16010
1252
28627
2260
44268
3236
63397
0.8 MPa
11.6(12.3)
225
27.1
696
53.8
1662
89
2753
171
5314
300
9276
761
20315
1410
36325
2546
56172
3646
80445
1 MPa
12.8
272
30
843
59
2012
98
3333
189
6435
330
11232
922
24595
1649
43976
2801
68005
4012
97390
1.5 MPa
15.3
390
36
1209
72
2884
119
4778
231
9224
402
16102
1322
35258
2364
63043
3655
97489
5235
139614
2 MPa
19.1
508
41
1575
93
3759
156
6228
300
12024
524
20986
1723
45953
3081
82165
4764
127058
6823
181960
2.5 MPa
23.6
628
49
1945
116
4640
192
7688
371
14842
647
25907
2127
56729
3803
101433
5882
156854
8423
224633
3 MPa
28.1
748
58
2318
138
5532
229
9166
442
17694
772
30883
2536
67624
4534
120913
7011
186978
10041
267772
(1) Maximum flow rate is the lower of 80 m/s.
(3) Minimum values are determined from Table 12. The values in parenthesis show the minimum linear flow rates (Re = 20,000 or 40,000)
when they are higher than the minimum measurable flow rate.
DYF Tab-16
min.
max.
min.
max.
min.
max.
min.
max.
min.
max.
min.
max.
min.
max.
min.
max.
min.
max.
min.
max.
DY015
DY025
DY040
DY050
DY080
DY100
DY150
DY200
DY250
DY300
DY025
-/R1
DY040
-/R1
DY050
-/R1
DY080
-/R1
DY100
-/R1
DY150
-/R1
DY200
-/R1
—
—
—
DY040
-/R2
DY050
-/R2
DY080
-/R2
DY100
-/R2
DY150
-/R2
DY200
-/R2
—
—
—
—
Model Code
Flow
Rate
Limits