9-14
IM 01F06A00-01E
9. GENERAL DESCRIPTION
Code
Specification
Item
T090408-1.EPS
SAA Flame proof Approval
Applicable Standard: AS 2380.1-1989, AS2380.2-1991, AS1939-1990
Type of protection: Ex d IIC T6...T1, IP67, Class 1, Zone 1
Amb.Temp.: –29 to +60
°
C (Integral Type Flowmeter and Remote Type Flowmeter)
: –40 to +60
°
C (Remote Type Converter)
Max. process temp. : T6; 85
°
C, T5; 100
°
C, T4; 135
°
C, T3; 200
°
C, T2; 300
°
C, T1;450
°
C
Electrical connection: ANSI 1/2 NPT female, ISO M20 X 1.5 female
SAA intrinsically Safe Approval (Note 2)
SAA Type n Approval
Applicable Standard: AS 2380.1-1989, AS2380.7-1987, AS2380.9-1991, AS1939-1990
Type of Protection: Ex ia IIC T4 IP67 (Integral Type Flowmeter, Remote Type Flowmeter and Remote
Type Converter)
Maximum Input Voltage (Ui)=30Vdc
Maximum Input Current (Ii)=165mAdc
Maximum Input Power (Pi)=0.9W
Internal Capacitance (Ci)=37nF
Internal Inductance (Li)=0mH
Ambient Temperature: –20 to +60
°
C
Type of Protection: Ex n IIC T4 IP67 (Integral Type Flowmeter, Remote Type Flowmeter and Remote
Type Converter)
Hazardous Area: Class I, Zone 2
Maximum Input Voltage (Ui)=30Vdc
Ambient Temperature: –20 to +80
°
C
Electrical connection: ANSI 1/2 NPT female, ISO M20
1.5 female
Standards
Association
of Australia (SAA)
(Note3)
Standards
Association
of Australia (SAA)
SS1
SF1
(Note 1) The flameproof packing adapter (/G11,G12) is necessary except the electrical conduit work. In case the ambient temperature
exceeds 50deg.C, use heat resistant cables with maximum allowable temperature of 70deg.C or above.
(Note 2) For intrinsically safe approval, use the barrier certified by the testing laboratories (BARD-400 is not applicable).
(Note 3) SAA Flame proof Approval (/SF1) can not combine with High Process Temperature Version (/HT).