9-9
IM 01F06A00-01E
9. GENERAL DESCRIPTION
Item
Specification
Code
(Note 1) The specified Tag Number is engraved on the data plate and stainless tag plate. The limitation of characters for Tag Number is, for BRAIN
Terminal or name plate, stainless steel tag plate: 16 characters, and for HART Configuration Tool: 8 characters.
(Note 2) There is a case that calibration water should stay in the meter tube. So this is not degrease treatment in the strict sense.
(Note 3) The output is set 3.6mA or less (General type is set 21.6mA or more at shipping).
(Note 4) The electrical connection turn to a downstream side.
(Note 5) Refer to “OPTION MULTI-VARIABLE (BUILD IN TEMPERATURE SENSOR) TYPE (/MV)” (see 9.4.2)
In case of Remote type detector (DY***-N), select “/MV” both DY and DYA.
(Note 6) /NM can not combine with Remote type (DY***-N).
(Note 7) SAA Flame proof Appreval (/SF1) can not combine with High Process Temperature Version (/HT).
(Note 8) • Cryogenic version (/LT) is not available.
• High process temperature version (/HT) and Multi-variable type (/MV) for DY025/R1 and DY040/R2 are not available.
• Explosion protected type SAA (/SF1,/SS1) is not available.
• Flange type only and available process connections are JIS10k, 20k (BJ1, BJ2) and ANSI150, 300 (BA1,BA2,BS1,BS2).
• Model Code size means “DY***-” nominal size.
(Note 9) • Applicable for option code /FF1, /FS1, /KF1, /KS1 and /KN1.
• Not applicable for option code: /P1, /P2, /P7, /X1, /X2, /HT, /LT, /SB, /JF3, /CF1, /CS1, /CF11, /CS11, /SF1, /SS1.
• The materials of exterior parts, name plate, screw, bolts on the stainless steel housing and bracket, u-bolt, nuts for DYA/E1 and tag plate for
/E1/SCT are JIS SUS316 or SUS316L.
Applicable Model
T090402.EPS
Each certificate to be attached produced by the vendors.
1. Meterbody
1. Meterbody, 2. Shedder bar
1. Meterbody, 2. Shedder bar, 3. Bottom plug
1. Meterbody, 2. Shedder bar, 3. Bottom plug, 4. Welding rod
Each certificate to beattached produced by the vendors.
1. Meterbody
1. Meterbody, 2. Shedder bar
1. Meterbody, 2. Shedder bar, 3. Bottom plug
1. Meterbody, 2. Shedder bar, 3. Bottom plug, 4. Welding rod
Positive Material Identification certificate to be attached for the main 3 chemical
components of specified materials. Each certificate to be attached.
1. Meterbody
1. Meterbody, 2. Shedder bar
1. Welder/Welding Operator Performance Qualification
(or Welder Qualification Record)
2. Welding Procedure Specification (WPS)
3. Procedure Qualification Record (PQR)
Each certificate to be submitted.
The customer’s name and job name to be specified when ordered.
1. Welded portion for the bottom plug
2. Welded portion for the flange in case of the welding
construction
Material certificates: Mill sheets
Material certificates: EN10204 3.1
PAMI test certificate
ASME welding documents submission
Liquid Penetrant test certificate
Item to be
specified
Item to be
specified
Item to be
specified
Item to be
specified
Item to be
specified
M01
M02
M03
M04
E01
E02
E03
E04
PM1
PM2
WP
PT
DY
DY
DY
DY
DY
1. Welded portion for the bottom plug
2. Welded portion for the flange in case of the welding
construction
Dye Penetrant test certificate for the welded portion to be attached.
Each certificate to be attached.
DY
2. is for DY250
and DY300.
Table 9.4.1 Paint Color and Codes
Codes
P1
Munsell Renotation Code
N1.5
Color
Black
P2
7.5BG4/1.5
Jade green
P7
Metallic silver
T090405.EPS
Table 9.4.2 Test Pressure Value
Pressure
2.1 MPa
5.0 MPa
2.9 MPa
Flange Rating
JIS 10 K
JIS 20 K
ANSI Class 150
7.5 MPa
ANSI Class 300
10.0 MPa
JIS 40 K
14.9 MPa
ANSI Class 600
22.4 MPa
ANSI Class 900
1.5 MPa
3.8 MPa
DIN PN 10
DIN PN 25
5.9 MPa
DIN PN 40
2.4 MPa
DIN PN 16
T090406.EPS
–29
+100
+200
+300
+450
+85
+60
–29
Fluid temperature (˚C)
Ambient temperature (˚C)
Operating range
F090401.EPS
Figure 9.4.1 Fluid temperature range of high process
temperature version
–196
–100
–29
0
0
–50
–20
Fluid temperature (˚C)
Ambient temperature (˚C)
Operating range
+100
F090402.EPS
–29
Figure 9.4.2 Fluid temperature range of cryogenic version