![YOKOGAWA DO202G User Manual Download Page 66](http://html1.mh-extra.com/html/yokogawa/do202g/do202g_user-manual_3499821066.webp)
IM 12J05C01-01E
Appendix 2-1
APPENDIX 2. QUALITY INSPECTION
A-2-1. DO202G 2-Wire Dissolved Oxygen Transmitter
Quality
Inspection
Standards
DO202G
Dissolved Oxygen Transmitter
QIS 12J05C01-01E
1st Edition: Feb. 2007
1. Scope
This inspection standard applies to the DO202G Dissolved Oxygen Transmitter.
2. Inspection
Items
2.1 Insulation resistance test
2.2 Current output test
2.3 Temperature indication check
* 2.4 Galvanic indication check
* 2.5 Polarograph indication check
Note: Items marked with an asterisk (*) may only be confirmed by a test certificate.
3.
Inspection Methods, Standards and Conditions
z
Connect the testing circuit as shown in Figure 1. Allow the instrument to warm up for at least
5 minutes before conducting the tests. For the connections for the insulation resistance test,
follow the instructions in Section 3.1.
z
Performance tests should be done in the inspection mode where the tests from Section 3.2
through Section 3.5 take place in sequence and cannot be retraced. If the reconfirmation of
a test is needed, turn off the power to the transmitter, turn on the power again, and enter the
inspection mode to restart the tests.
3.1 Insulation Resistance Test
(1) Apply 500 V DC between the power supply terminals shorted together (+ and –) and the
earth terminal (G). The insulation resistance must be 100 M
Ω
or greater.
(2) Apply 500 V DC between the input terminals shorted together (11 to 18) and the earth
terminal (G). The insulation resistance must be 100 M
Ω
or greater.
3.2 Current Output Test
Connect the testing circuit as shown Figure 1 and set the equipment as follows:
Decade resistance box: 1019.5 [
Ω
]
DC generator 1:
0
μ
A
DC generator 2:
0
μ
A
DC source:
24 V DC
User the following procedure to enter the inspection mode.
(1) Entering Service Code 87.
a. Press the [
] key. The message display will show “
OUTP.”
b. Press the [NO] key twice. The message display will show “
SERV.”
c. Press the [YES] key. The data display will show “00” with the first digit of 0 flashing.
d. Press the [
] key eight times. The data display will show “80” with the first digit of 8
flashing.
e. Press the [>] key once. The data display will show “80” with the second digit of 0
flashing.
f. Press
the
[
] key seven times. The data display will show “87” with the second digit of 7
flashing.
g. Press the [ENT] key. The message display will show “
PASS
.”
Summary of Contents for DO202G
Page 7: ...IM 12J05C01 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 11: ...IM 12J05C01 01E 1 4 Introduction...
Page 31: ...IM 12J05C01 01E General Specifications 2 20...
Page 37: ...IM 12J05C01 01E...
Page 65: ...IM 12J05C01 01E...
Page 69: ...IM 12J05C01 01E 2 4 Appendix...
Page 74: ...IM 12J05C01 01E Appendix 2 9...
Page 78: ...IM 12J05C01 01E Appendix 2 13...
Page 82: ...IM 12J05C01 01E Appendix 2 17...