IM 12J05C01-01E
Installation and Wiring 3-4
3-3-2. Wiring other galvanic sensors
Consult the users manual for the color identification
of the sensor cable.
- The temperature sensor has two wires and must
be connected to the terminal 11 and 12.
- The measuring electrode: the cathode must be
connected to terminal 13.
- The reference electrode: the anode must be
connected to terminal 15.
- The overall shield of the cable must be connected
to terminal 14 if there is one available.
- Place the jumper in the GALVANIC position.
3-3-3. Wiring polarographic sensors
Consult the owners manual for the color identifica-
tion of the sensor cable.
- The temperature sensor has two wires and must
be connected to the terminal 11 and 12.
- The measuring electrode: the cathode must be
connected to terminal 17.
- The reference electrode: the anode must be
connected to terminal 18.
- The overall shield of the cable must be connected
to terminal 14 if there is one available.
- Place the jumper in the POLAROGRAPHIC position.
3-3-4. Wiring sensors with Vp connector
Consult the owners manual for the color identifica-
tion of the sensor cable
- The temperature sensor has two wires and must
be connected to terminal 11 and 12
- Polarographic : the measuring electrode
(cathode) must be connected to terminal 17
- Polarographic : the reference electrode (anode)
must be connected to terminal 18
- Galvanic : the measuring electrode (cathode)
must be connected to terminal 13
- Galvanic : the reference electrode (anode) must
be connected to terminal 15
- The overall shield of the cable must be connected
to 14
- Place the jumper in the Polarographic or Galvanic
position
Figure 3-6. Connections DO
Figure 3-7. Terminal indentification labels
A-17 pol. 13 galv.
B-18 pol. 15 galv.
C
D
E-11
F-12
S-14
temp
11
12
12
13
14
14
16
15
13
14
14
16
15
17
11
17
11 Red
12 Blue
15 Core 16 Screen
White Co-axial cable
13 Core 17 Screen
Brown Co-axial Cable
WF10 Cable
14 Overall Screen
TRANSMITTER / CONVERTER
11
12
17
13
15
16
14
A
C
Overall
shield
B
D
Screen
E
Red
White
Blue
Brown
Figure 3-8. Connection of WF10 extension cable and BA10/BP10 junction box
3-3-5. Sensor cable connections using junction
box (BA10) and extension cable (WF10)
Where a convenient installation is not possible using
the standard cables between sensors and trans-
mitter, a junction box and extension cable may be
used. The Yokogawa BA10 junction box and the
WF10 extension cable should be used. These items
are manufactured to a very high standard and are
necessary to ensure that the specifications of the
system are not compromised. The total cable length
should not exceed 30 metres (including DO sensor
cable).
Summary of Contents for DO202G
Page 7: ...IM 12J05C01 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 11: ...IM 12J05C01 01E 1 4 Introduction...
Page 31: ...IM 12J05C01 01E General Specifications 2 20...
Page 37: ...IM 12J05C01 01E...
Page 65: ...IM 12J05C01 01E...
Page 69: ...IM 12J05C01 01E 2 4 Appendix...
Page 74: ...IM 12J05C01 01E Appendix 2 9...
Page 78: ...IM 12J05C01 01E Appendix 2 13...
Page 82: ...IM 12J05C01 01E Appendix 2 17...