IM 12J05C01-01E
Error Generation
Error content and Causes
Measures
No. Mode
E10 All modes
EEPROM writing failure
Turn off the power and then turn it
Electronic circuit failure
back on and check whether or not the
system returns to normal.
If the failure occurs again, contact
Yokogawa to request repair.
E15 CODE 11
Temperature error correction failure
Accurately measure the temperature of
The difference between the corrected value
the measuring solution (dissolved oxygen
and the standard value exceeds the range
sensor).
corresponding to ±7.5°C (±13 5°F).
Examine whether the sensor cable fails
• Temperature entry is not accurate.
(partial disconnection or degraded
• Sensor cable failure
insulation) and then correct the
temperature error again.
E16 All modes
Call for maintenance-interval time exceeded
Perform maintenance. Reset interval.
E17 *OUTP
Output span too small.
Re-enter a value within the permissible range.
CODE 35
Min. span is 1ppm, 25ppb or 10%
E18 CODE 35
Table values make no sense.
Reprogram.
E19 At data entry
Configuration failure
Re-enter a value within the permissible
A value not within the permissible range is entered.
range.
E20 All modes
Instrument initialization failure
Contact Yokogawa
E21 All modes
EPROM comparison failure
Contact Yokogawa
Troubleshooting 8-2
Note:
It is possible that the error message does
not disappear immediately after corrective
measures.
Then turn power off and the turn it back on
and check whether the system returns to
normal.
mA
mA
Summary of Contents for DO202G
Page 7: ...IM 12J05C01 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 11: ...IM 12J05C01 01E 1 4 Introduction...
Page 31: ...IM 12J05C01 01E General Specifications 2 20...
Page 37: ...IM 12J05C01 01E...
Page 65: ...IM 12J05C01 01E...
Page 69: ...IM 12J05C01 01E 2 4 Appendix...
Page 74: ...IM 12J05C01 01E Appendix 2 9...
Page 78: ...IM 12J05C01 01E Appendix 2 13...
Page 82: ...IM 12J05C01 01E Appendix 2 17...