IM 12J05C01-01E
5-9 Parameter Setting
Code Display
Function
Function
detail
X Y Z Default
values
Output Functions
30
Not
used
31
*OUTP.F
mA output functions
Linear
0
0
Linear
Table
1
32
*BURN
Burn function
No burnout
0
0
No Burn.
Burnout
downscale
1
Burnout
upscale
2
Pulse
burnout
3
33, 34
Not used
35
*TABLE
Output table for mA
*0%
Linearization table for mA in 5% steps.
*5%
The measured value is set in the main
*10%
display using the >, ^, ENT keys, for
...
each of the 5% interval steps.
...
Where a value is not known, that value may
*95%
be skipped, and a linear interpolation will
*100%
take
place.
36-39
Not
used
Code Display Function
Function
detail
X
Y
Z
Default
values
User Interface
50
*RET
Auto return
No return to meas. from HOLD
0
1
Return to meas after 10 min
1
51
*P.MAN
Manual pressure No manual pressure in Maint.mode
0
0
Manual pressure in Maint.mode
1
*P.%
Enter Manual process pressure 0-999.9 kPa
101.3 kPa
*P.CAL
Enter Cal. pressure in Maint.mode 0-999.9 kPa
101.3 kPa
52
*PASS
Password
Direct access to MAINT
0
0.0.0
Pass: 111 for #=1, MAINT protected by password
#
333 for #=2,
Direct access to COMM
0
777 for #=3,
COMM protected by password
#
888 for #=4,
Direct access to SERV
0
123 for #=5,
SERV protected by password
#
957 for #=6,
331 for #=7,
546 for #=8 and
847 for #=9
53
*Err.01
Error setting
Calibration not stable
Soft fail
0
Calibration not stable
Hard fail
1
1
Hard
*Err.02
Zero out of limits
Soft fail
0
Zero out of limits
Hard fail
1
1
Hard
*Err.07
Temperature sensor open
Soft fail
0
Temperature sensor open
Hard fail
1
1
Hard
*Err.08
Temp. sensor shorted
Soft fail
0
Temp. sensor shorted
Hard fail
1
1
Hard
*Err.09
Cell current abnormal
Soft fail
0
Cell current abnormal
Hard fail
1
1
Hard
*Err.16
Call for maintenance
Soft fail
0
0
Soft
Interval time exceeded
Call for maintenance
Hard fail
1
Interval time exceeded
55
*CALL.M E16 setting
No call for maintenance
0
0
Call for maint. enabled
1
*CALL .M
E16 message after 1- 250 days
250
56
*UNIT
PV unit
Unit of measurement ppm
0
0
Unit of measurement ppb
1
Unit of measurement % sat.
2
57-59
Not
used
mA
Summary of Contents for DO202G
Page 7: ...IM 12J05C01 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 11: ...IM 12J05C01 01E 1 4 Introduction...
Page 31: ...IM 12J05C01 01E General Specifications 2 20...
Page 37: ...IM 12J05C01 01E...
Page 65: ...IM 12J05C01 01E...
Page 69: ...IM 12J05C01 01E 2 4 Appendix...
Page 74: ...IM 12J05C01 01E Appendix 2 9...
Page 78: ...IM 12J05C01 01E Appendix 2 13...
Page 82: ...IM 12J05C01 01E Appendix 2 17...