IM 12J05C01-01E
Appendix 1-1
A-1-1. User setting table
FUNCTION
SETTING DEFAULTS
USER SETTINGS
Parameter specific functions
01 *S.TYPE 1
Polarographic
*V.POL 0.675 V
*SENS 7.5
nA/ppm
04 *SAL.TY 0
Off
*[Cl-] 0
g/kg
Temperature measuring functions
10 *T.SENS 1
22k
NTC
11 *T.UNIT 0
°C
12 *T.ADJ
None
13 *T.MAN 0
Off
Calibration parameter functions
20 *
∆
T.SEC 60
sec.
*
∆
Z.PPM 0.1
ppm
*
∆
Z.PPB 1
ppb
*
∆
Z.SAT 1.0
%
*
∆
S.PPM 0.1
ppm
21 *
∆
Z.CAL 0
Off
*
∆
Z.LIM 199.9 nA
(polar)
22 *ZERO
0
μ
A (nA)
*SENS 7,5
nA/ppm
mA outputs
31 *OUTP.F 0
Linear
ppm
32 *BURN
0
No
Burn
35
*TABLE
0 to 20
ppm
User Interface
50 *RET
1
on
51 *P.MAN 0
off
*P.% 101.3 kPa
*P.CAL 101.3 kPa
52 *PASS
0.0.0
all
off
53 *Err.01
1
hard
fail
*Err.02 1
hard
fail
*Err.07 1
hard
fail
*Err.08 1
hard
fail
*Err.09 1
hard
fail
*Err.16 0
soft
fail
55 *CALL.M 0
Off
*CALL.M 250
days
56 *UNIT
0
ppm
Communication
60 *COMM. 1.0
on/write
enable
*ADDR. 00
61 *HOUR
62 *ERASE
General
70 *LOAD
Test and setup mode
80 *TEST
mA
mA
mA
APPENDIX 1. USER SETTINGS
Summary of Contents for DO202G
Page 7: ...IM 12J05C01 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 11: ...IM 12J05C01 01E 1 4 Introduction...
Page 31: ...IM 12J05C01 01E General Specifications 2 20...
Page 37: ...IM 12J05C01 01E...
Page 65: ...IM 12J05C01 01E...
Page 69: ...IM 12J05C01 01E 2 4 Appendix...
Page 74: ...IM 12J05C01 01E Appendix 2 9...
Page 78: ...IM 12J05C01 01E Appendix 2 13...
Page 82: ...IM 12J05C01 01E Appendix 2 17...