![YOKOGAWA AQ1210D User Manual Download Page 285](http://html1.mh-extra.com/html/yokogawa/aq1210d/aq1210d_user-manual_3572414285.webp)
11-15
IM AQ1210-01EN
Troubleshooting, Updating, and Storage
11
22
33
44
55
66
77
88
99
10
10
11
11
App
App
Index
Index
11.8 Routine Maintenance
1.
Make sure that the instrument is turned off.
2.
Open the cover of the OTDR port or light source port (option).
3.
Remove the optical adapter.
For instructions, see “Replacing the Optical Adapter” in the Getting Started Guide, IM AQ1210-02EN.
4.
Use a lens cleaner to clean the optical fiber end face.
5.
Insert the lens cleaner slowly into the optical adapter’s optical connector and turn the lens
cleaner to clean the inner walls.
Lens cleaner
Optical adapter
Lens cleaner
Optical fiber end face
Optical connector
OTDR port cover
Guide hole
Note
• Use a lens cleaner that does not produce fuzz or residue to clean the optical components.
• The OTDR port or light source port can be cleaned with a stick type lens cleaner with the optical adapter
attached, but we recommend that you remove the optical adapter to clean it.