ML-x17 DATA LOGGER MANUAL
Manufacturers of low power instruments
page 22
Now we can setup the FTP, TCP and Email output
Therefore, go to menu-option D: <Modem output> and choose the wished output protocol e.g. Email
It should look like this:
Email settings
Beware:
Some providers
strictly check
the Originator
address. So
make sure this
address is valid.
Now perform an Email test and check if it is working right.
It is strongly recommended to include some internal sensors in the configuration, because of the
monitoring of the performance of the data logger itself. Most users like to keep track of the battery-life for
example.
Go To menu-option 6 <internal sensors>
[0] Exit
[1] Name >> Email
[2] Send interval >> Normal 01:00:00; Alarm Not Used
[3] Send delay >> Not Used
[4] SMTP Server >> smtp.provider.com
[5] SMTP Port >> 25
[6] User name >> reh6a01
[7] Password >> ********
[8] Originator address >> [email protected]
[9] Destination address >> [email protected]
[A] Subject >> YDOC upper stream
[R] Remove
[T] EMAIL test >> Not Done
>
Internal sensors
[0] Exit
[1] Name >> Internal
[2] Sample interval >> Data log interval
[3] Battery capacity >> 14000 mAh
[4] Battery replaced >>
Yes
[5] Rest capacity >> Not used
[6] Rest power >> Not used
[7] Processor temperature >> Processor
temperature
[8] Average voltage >> Not used
[9] Max voltage >> Not used
[A] Min voltage >> Min voltage
[B] Average current >> Not used
[C] Max current >> Not used
[D] Min current >> Not used
[E] Operating cycle >> Operating cycle
[F] Free disk space >> Not
used
[R] Remove
[T] Test measurement
>