ML-x17 DATA LOGGER MANUAL
Manufacturers of low power instruments
page 121
First, change the name “Digital Pulse” into a more comprehensive one. We use “Rain” here.
The menu-
item “pulses per unit” is very important and converts the input pulses into a physical value.
It is advised to test the hardware first, before proceeding to selecting the right settings for bucket-size etc.
Therefore, leave this value (1) and test your sensor first.
To test, just connect it, and apply a known amount of pulses tot the data logger. You can verify this count
with the command
<Ctrl>A<Shift>V<Ctrl>D
When this is correct you can proceed to set up your
rain-gauge.
Now you have to enter the physical details of your rain-gauge.
Digital pulse sensor
[0] Exit
[1] Name >> Rain
[2] Sample interval >> Data log
interval
[3] Port mode >> Port 1; Internal pull up
[4] Register mode >> Pulse (low frequency)
[5] Units per pulse >> 0.2
[6] Register value >> 16979 pulses
[7] Register reset >> Off
[8] Log each counter change >> On
[9] Counter (unit) >> Counter
[A] Quantity (unit) >>
Quantity
[B] Mean rate (unit/h) >> Mean Rate
[C] Max rate (unit/h) >> Max
Rate
[D] Min rate (unit/h) >> Min Rate
[R] Remove
>