![Yamato WG 1000 Instruction Manual Download Page 42](http://html.mh-extra.com/html/yamato/wg-1000/wg-1000_instruction-manual_905422042.webp)
39
Long storage and disposal
When not using this unit for long term / When disposing
•
If this unit is to be put out of service for a long time, be sure to turn of the earth leakage
breaker of this unit for safety, and close the tap. Water in the boiler and distilled water tank, if
stored as it is, will deteriorate in quality due to generated bacteria or algae. Drain water by
the procedure below:
Boiler Water Drain
1. In draining boiler water, turn off the earth
leakage breaker, ensure that the tap is
closed, then wait for 30 minutes or more,
and open the front door.
2. Open the boiler water drain cock.
3. Make sure that all water in the boiler and
float cylinder is drained.
4. Be sure to close the boiler water drain
cock. If boiler water drain cock is opened
in next use, water is not fed into the boiler,
and distillation is not started.
Boiler Water
Drain Cock
Open
Close