![Yamato WG 1000 Instruction Manual Download Page 39](http://html.mh-extra.com/html/yamato/wg-1000/wg-1000_instruction-manual_905422039.webp)
36
Maintenance Method
Washing of Distiller
2. Attach 4 heater lead terminals to the terminal block.
Heater Lead
(wire diameter, thick)
Heater Lead
(wire diameter, thin)
3. Install the left side plate on the body.
4. Insert the hose to the boiler water supply and drain port, and secure with the hose band.
Installation of condenser
1. Place packing in the connection port of boiler with condenser, and secure with connecting hardware
so that the boiler and condenser are placed in the same direction.
Inlet
Outlet
Boiler
Condenser
Connecting
Hardware
2. Connect the hose respectively to the cooling water inlet, outlet, and distilled water outlet of
condenser.
3. Close the boiler water drain cock.
View from above
Boiler Water
Drain Cock
Open
Close