18
4. Operation procedures
Operation Method-Spaying Operation
(11) After the outlet temperature is stabled, please
change the sample to actual samples. At this
moment, the outlet temperature will be changed,
adjust the infuse speed if necessary.
~Finished operation~
(12) After the sample infusion is finished, add
distilled water to clean the nozzle. Switch the
pump FWD switch to OFF, adjust the spraying
pressure to 0.
(13) Turn OFF the heater, when the inlet temperature
is below 45
°
C, switch of the wind volume. Turn
OFF the Blower.
To avoid any trouble, please do not stop the
blower when the outlet temperature is above
45
°
C
.
(14) Please turn off the power switch.
(15) Untie the container fasten ribbon, take out product
catcher. At this moment, please note there is also
powder attached inside the cyclone cover.
(16) Please follow the cleaning method (P24.
「
On
clearance after use
」
to clean up the container)
When using the metal corrosive samples such
as sodium chloride, please be sure to clean it
carefully after disassembling.
Heater on after blower knob set
Be sprayed at Stable Temp.
Set Spray Pressure & Flow