Installation
MIQ/R6
14
ba75541d07
03/2018
Connecting lines
to the terminal
strip
Fig. 3-1
Terminal strip with the relay connections
1
Open the module.
2
Screw the cable gland (pos. 1 in Fig. 3-1) with the sealing ring (pos. 2)
into the module housing.
3
Loosen the coupling ring (pos. 3 in Fig. 3-1).
4
Feed the line through the cable gland in the module housing.
5
Connect the wires to the terminal strip. While doing so, pay attention to
the specifications on the label located under the terminal strip.
6
Tighten the coupling ring (pos. 3 in Fig. 3-1).
WARNING
No free wires are allowed to project into the housing. Oth-
erwise, there is a danger that areas safe to contact could
come into contact with dangerous voltages. Always cut
off any wires that are not in use as closely as possible to
the cable gland.
1
3
2
X6
X5
X4
SENSORNET 2
RED
RO
T
SHIELD
SCHIRM
GREEN
GRÜN
X3
X2
X1
SENSORNET 1
RED
RO
T
SHIELD
SCHIRM
GREEN
GRÜN
ON
EIN
OFF
AUS
SN
TERMINA
TO
R
2
2
Relay output
X18 X17
240 VAC
2 A AC
R6
X16 X15
240 VAC
2 A AC
R5
X14 X13
240 VAC
2 A AC
R4
X12
X10
X8
X11
X9
X7
240 VAC
2 A AC
R3
240 VAC
2 A AC
R2
240 VAC
2 A AC
R1
Segment 2
Segment 1