Installation
MIQ/R6
10
ba75541d07
03/2018
– Temperature resistance (minimum and maximum temperature)
– Stability against UV light when used outdoors
– Protection against water and dust (Nema or IP protection class).
Suitable fusing of the electrical circuit
– Overcurrent fuses (in accordance with the technical data of the device
input or output)
– Overvoltage class II surge limiters
Suitable disconnecting device (e. g. switch or circuit breaker) for the mains
supply of permanently mounted equipment with separate mains connection
– in compliance with the following regulations
- IEC 60947-1
- IEC 60947-3
– in the vicinity of the equipment (recommendation)
Flame-resistant (cables and lines),
in compliance with the following regulations
– UL 2556 VW-1 (for USA, Canada)
– IEC 60332-1-2 (outside of USA, Canada)
3.4
Installation in the IQ S
ENSOR
N
ET
The IQ S
ENSOR
N
ET
provides a number of options for integrating the MIQ/R6
mechanically and electrically in the system (stacked mounting, distributed
mounting, etc.). The various types of installation are described in detail in the
I
NSTALLATION
chapter of the system operating manual.