7.1.1 Error messages
7.1.2 Info messages
The MIQ/R6 output module does not send any info messages.
Information about
Contents and structure of the log book and
Structure of the message code
can be found in the L
OG
BOOK
chapter of the IQ S
ENSOR
N
ET
system
operating manual.
All message codes of the MIQ/R6 output module end with the num-
ber "431".
Message code
Message text
EI2431
Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
*
Check terminal connections and module connections
* Defective components,
replace components