pHix® Compact V3. NG (next generation)
EN 4.10 Version 2202
pHix Configurator software /Register list
86
Liquid ground
(Potential equalization)
Internal
External
Internal is only for the PPS sensor
version
External is only for the stainless steel
version
Switch ring
No switch ring
Reed 4
Reed 7
Reed 10
Reed M
Disabled
Switch ring is only for pHix 4.6 pH
and 7.0 pH
Show the ring position of the sensor.
adjustment using the switch ring
Tilt
Close/Up
Open/Down
Disabled
Tilt is only for pHix 4.6 pH and 7.0 pH
Show the position of the sensor.
adjustment using the tilt switch
Register list
General descriptions
Modbus protocol is used to read/write registers.
Read registers :
Function code 3: Read Multiple Holding Registers
Write registers :
Function code 16: Write Multiple Holding Registers
Data types
The following data types are used in pHix units, when reading registers from a register
list.
Name
Description
U16
Unsigned word (16 bit) [WORD]
F32
Float (32 bit) comply with IEEE754 [FLOAT