pHix® Compact V3. NG (next generation)
EN 4.10 Version 2202
pHix Configurator software /Main configuration
60
Menu item Option
Description
Device
type
pHix pH
The sensor type
Electrode
type
pH
pH-
Electrode
zero type
Zero
point 4.6
pH
Zero
point 7.0
pH
The pH electrode type
Serial
number
The sensor serial number
Part
number
The sensor item number
Sensor on-
chip serial
number
Chip serial number
Number &
Revision
The current firmware in the sensor
Compile
time
The last time the firmware was changed
Slope
The slope is determined by measuring the difference in mV
readings pH buffers. The ideal slope should be approximately
-59.16 mV/pH at 25°C
Zero point
The pH where the Zero Point is 0 mV, it means the same
amount of H+ and OH- ions.
pHix Redox