pHix® Compact V3. NG (next generation)
EN 4.10 Version 2202
Mounting /Eletrical mounting
19
Eletrical mounting
pHix® Compact must not be installed in explosion hazardous locations!
pHix®
Compact should be connected to an active 4-20 mA input or supplied from a separate
10-28 VDC voltage source. Max power consumption is 50 mW.
pHix® Compact 4-20 mA output(s) is galvanically separated from the liquids potential.
Signal cable
pHix® Compact comes with two galvanically separated mA signal outputs and Modbus
RS485.
Note!
pHix® Compact is delivered with cable made of PUR.
pHix® Compact with 2 x 4-20 mA outputs, one for pH or redox measurement and one
for temperature measurement.