pHix® Compact V3. NG (next generation)
EN 4.10 Version 2202
Mounting /Connection examples
20
Cutting the signal cable
The cable comes with stripped wire ends ready for mounting.
pHix® Compact uncut cable-end with two signal outputs.
Extending the signal cable
There are no specific demands to the cable that may be used
for extending the existing cable, except that it should be
shielded and that the voltage drop should not be too big.
Note!
pHix® Compact is a passive transmitter. Considerations
should be taken regarding the voltage drop over the signal
cable.
The supply voltage on the site must not drop below 10 V DC at
maximum current (20 mA.)
In order not to compromise operational reliability, the cables
should be interconnected by means of a watertight connection
box (item no. 200590).
MJK connection box (item no. 200590).
Connection examples
In order to communicate by RS-485, the pH loop must be applied by 10-28VDC.