June 2022 – TD 320 OPERATING MANUAL – TURBIDITY SENSOR 4296
7.1.1 Function test connected to SEAGUARD
®
with display
Leave the Turbidity Sensor 4296 mounted onto the
SEAGUARD
®
Platform. Power the instrument; refer TD262a
and TD262b for operating instructions.
Open
Administrative Tools-> Sensor Monitor
Sensor Monitor
can be used as a direct reading of the
sensor; the function is mainly used for test purposes.
Select the Turbidity Sensor from the list and press
Start
,
refer
Figure 7-2
.
The next window shows sensor information like the
Node Description, Product Name and Number,
and
Serial Number
, refer
Figure 7-3
.
The number of sensor parameters and the processing time
can be viewed in the window.
Select an
Update Interval
for the sensor monitoring.
Press
Start
to start monitoring the sensor readings.
Figure 7-2: Select sensors to monitor
Figure 7-3: Set the Update Interval